Accuracy and Reliability of Intermittent Scanning and Real-Time Continuous Glucose Monitoring Systems in Diabetes
Lukas van Baal1, Lutz Heinemann2, Johanna Reinold3
1Department of Endocrinology, Diabetes and Metabolism, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
Background:
Diabetes care is a major challenge of patients treated in hospitals. A continuous glucose monitoring system (CGM) provides a more comprehensive assessment of glucose control than capillary blood glucose measurements. Especially in emergencies, data on CGM use in inpatients are limited. To evaluate real-world usability, accuracy of an intermittent scanning and a real-time CGM in patients admitted due to diabetes emergencies was assessed.
Methods:
In 151 patients admitted due to diabetes emergencies, this single-center prospective study investigated the mean absolute relative difference (MARD) in broad glycemic ranges. The CGM accuracy was evaluated by applying a modified version of the Food and Drug Administration (FDA) criteria for CGM use, Clark Error Grid (CEG), and Bland Altman analysis (BAA).
Results:
Analysis of 1,498 CGM-/POC-glucose (CGM-/POC-G) pairs revealed a MARD of 10.8% with stepwise improvement from the hypoglycemic to the hyperglycemic range. The CEG analysis showed that 99.1% of all glucose values fell within the optimal or acceptable zones. BAA indicated that 96.0% of CGM-G values fell within the limits of the POC-G values. Day-by-day analysis of overall MARD revealed the highest MARD on the first day of CGM use, followed by consistent and stable MARD levels from day 2 onward until the end of wear time. Applying a modified version of the %20/20 agreement rate of the FDA criteria, 90.7% of CGM-G laid within ±20 mg/dl/±20% agreement rule.
Conclusion:
This study indicates the usability of CGM for inpatient diabetes care by demonstrating a high accuracy and reliability of CGM measurement.
Related Concept Videos
Hypoglycemia and Glucagon
Diabetes: Symptoms, Diagnosis, and Complications
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
Errors occurring during blood pressure monitoring
Several factors...
Diabetes Mellitus: Type 2 and Gestational
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...