Retinal Vascular Reactivity as Assessed by Optical Coherence Tomography Angiography
Published on: March 26, 2020
OPTICAL COHERENCE TOMOGRAPHY INTERFACE REFLECTION IMAGE ASSOCIATED WITH INTRAVITREAL GAS
1Prism Vision Group, South Jersey Eye Physicians Division, Moorestown, New Jersey; and.
Optical coherence tomography (OCT) interface reflection (IR) images can mimic macular holes after surgery. Awareness of this artifact, caused by fluid-gas interfaces, is crucial for accurate post-operative assessment.
Area of Science:
- Ophthalmology
- Medical Imaging
Background:
- Intravitreal gas injections are used in treating various retinal conditions.
- Accurate interpretation of post-operative optical coherence tomography (OCT) is essential for patient management.
Purpose of the Study:
- To describe and characterize an optical coherence tomography (OCT) interface reflection (IR) artifact.
- To illustrate how this artifact can simulate a full-thickness macular hole in a post-operative eye with intravitreal gas.
Main Methods:
- Retrospective case report analysis.
- Utilized spectral domain OCT (Cirrus HD-OCT 6000) imaging of a post-operative eye with intravitreal gas.
- Examined OCT images and slice navigator data.
Main Results:
- An OCT interface reflection (IR) image was observed anterior to the macula.
- This artifact mimicked a type 2 closure of a central full-thickness macular hole.
- Retrospective analysis confirmed the image was an artifact from a fluid-gas interface reflection, not the fovea.
Conclusions:
- OCT interface reflection (IR) artifacts can occur due to fluid-gas interfaces.
- These artifacts may be misinterpreted as macular pathology, potentially confusing clinical assessment.
- Clinicians must recognize these IR images to avoid misdiagnosis and ensure appropriate post-operative planning.