[Spatial Resolution and Uniformity of a Full-ring CZT SPECT/CT System: Comparison with a Conventional Anger-type
Takashi Takeuchi1, Yoshitaka Tanaka1,2, Yasuhiro Kodama1
1Department of Radiological Technology, Tohoku University Hospital.
Purpose:
StarGuide (GE HealthCare, Haifa, Israel) is a full-ring SPECT/CT system based on Cadmium Zinc Telluride (CZT) technology. In this study, we aimed to compare the image quality of this CZT-based SPECT/CT to a conventional Anger-type SPECT/CT system (NM/CT 870 DR, 870DR; GE HealthCare).
Methods:
Tomographic sensitivity was calculated by recording the total number of counts detected during tomographic acquisition for a point source. We evaluated spatial resolution and image uniformity on each system using the full width half maximum (FWHM) of line sources and root mean square uniformity (%RMSU) of pool phantom, respectively. The voxel size of the StarGuide SPECT images was 2.46×2.46×2.46 mm3, compared to 4.42×4.42×4.42 mm3 on 870DR. These projection data were reconstructed using 3D-OSEM with a resolution recovery technique (RR). We compared 3 different algorithms: non-correction (NCRR), scatter correction (SCRR), and attenuation correction and scatter correction (ACSCRR).
Results:
Tomographic sensitivity of StarGuide and 870DR were estimated at 200.0 counts・s-1・MBq-1 and 193.3 counts・s-1・MBq-1, respectively. Spatial resolution at the center of the FOV was estimated at 2.6 mm for StarGuide and 5.4 mm for 870DR with ACSCRR. Likewise, the %RMSU was 21.7 for StarGuide and 24.6 for 870DR.
Conclusion:
The full-ring CZT SPECT/CT system has a superior spatial resolution and better image uniformity than the conventional Anger-type SPECT instrument, whereas tomographic sensitivity remains similar.
More Related Videos
14:19A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Mass Analyzers: Overview
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....