Abstract:
The structural and dynamical properties of soft-matter systems play an important role in crystallization and nucleation theory. Despite their significance, the dynamical properties are still poorly understood because of experimental constraints and the requirement of performing measurements with high spatial and temporal resolution. Here, we demonstrate MHz X-ray photon correlation spectroscopy (XPCS) using a contactless sample holder at the European X-ray Free-Electron Laser. A millimetre-sized liquid sample droplet was levitated in air via acoustic waves with the solvent slowly evaporating. A colloidal suspension of silica nanospheres was used to track the structural evolutions using small-angle X-ray scattering, and the dynamical information was captured by time-resolved MHz XPCS as a function of evaporation time. This study outlines a new path towards the investigation of metastable structure and dynamics using X-ray speckle techniques, for instance, XPCS, X-ray speckle visibility spectroscopy and X-ray cross-correlation analysis.
More Related Videos
08:44Measurements of Long-range Electronic Correlations During Femtosecond Diffraction Experiments Performed on Nanocrystals of Buckminsterfullerene
Published on: August 22, 2017
10:12Synchrotron X-ray Microdiffraction and Fluorescence Imaging of Mineral and Rock Samples
Published on: June 19, 2018
Related Concept Videos
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal crystal...
X-ray Imaging
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.