Quantitative chest computed tomography: regional differences in dual-energy-derived virtual vs. true non-contrast
Q D Strotzer1, C Schachner2, L Scheuermeyer2
1Institute of Radiology, University of Regensburg Medical Center, Franz-Josef-Strauss-Allee 11, 93053 Regensburg, Germany; Division of Neuroradiology, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, 55 Fruit St, Boston, MA 02114, USA.
Aim:
Quantitative analysis of chest computed tomography (CT) usually requires non-contrast scans. Multi-energy CT-derived virtual non-contrast (VNC) images may obviate additional true non-contrast (TNC) scans. We aimed to quantitatively compare virtual and true non-contrast images, with a focus on regional differences in pulmonary tissue.
Materials And Methods:
This bi-center retrospective, IRB-approved study included consecutive patients who received contrast-enhanced dual-energy and non-contrast chest CT scans within one year between April 2018 and December 2022. Scans were co-registered and semi-manually segmented into various volumes of interest: whole lung, center/periphery, ventral/dorsal half and upper/middle/lower third. The mean lung density, high and low attenuation volumes and the lower 15th percentile were computed from the VNC and TNC scans for each volume. Correlation and agreement were assessed using Pearson's and intraclass correlation coefficients and Bland-Altman analysis. Subgroup analyses involved body mass index and markers of pulmonary emphysema.
Results:
Forty-eight patients were included (26 females, median age: 63 years, interquartile range: 58-69). Pearson's and intraclass correlation coefficients > 0.9 demonstrated high correlation and agreement for the mean lung density, low attenuation volume and the lower 15th percentile across the whole lung. The agreement for high attenuation volume was slightly less pronounced (intraclass correlation coefficient, 0.84 [95% confidence interval: 0.73, 0.91]). Regional differences were observed, for example, concerning the mean lung density with mean differences (HUVNC minus HUTNC) of -5.71 HU [-11.07, -0.35], 1.62 HU [-2.98, 6.23] and 5.77 HU [-0.47, 12.01] for the upper, middle and lower third, respectively.
Conclusion:
Although slight regional differences were observed, quantitative analysis can be reliably performed using VNC images.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...