Systematic comparison of Commercial seizure detection Software: Update equals Upgrade?
Johannes Koren1, Kady Colabrese2, Manfred Hartmann3
1Karl Landsteiner Institute for Clinical Epilepsy Research and Cognitive Neurology, Vienna, Austria; Department of Neurology, Clinic Hietzing, Vienna, Austria.
Encevis 2.0 demonstrated higher seizure detection rates, while Persyst 14 offered fewer false positives in this updated software comparison. Both seizure detection software packages are suitable for clinical use with proper expertise.
Area of Science:
- Neurology
- Medical Technology
- Epilepsy Research
Background:
- Commercial seizure detection software is crucial for epilepsy management.
- Previous versions of encevis and Persyst have been evaluated.
- Software updates necessitate re-evaluation of performance.
Purpose of the Study:
- To compare the performance of updated encevis 2.0 and Persyst 14 seizure detection software.
- To assess detection rates and false positive rates in a large video-EEG dataset.
Main Methods:
- A follow-up study evaluated two updated seizure detection software packages (encevis 2.0, Persyst 14).
- Data comprised 80 patients, 6826 hours of EEG, and 912 seizures.
- Different sensitivity settings were utilized for analysis.
Main Results:
- Encevis 2.0 achieved higher detection rates (80.7%-89.1%) compared to Persyst 14 (67.3%-71.1%).
- Persyst 14 exhibited significantly lower false positive rates (2-4 per day) than encevis 2.0 (12-22 per day).
- Mean seizure count was 11.4 per patient, with a 0.7 tonic-clonic seizure rate.
Conclusions:
- Encevis 2.0 offers superior sensitivity, while Persyst 14 provides better specificity.
- Updated software showed divergent performance changes: encevis improved sensitivity, Persyst improved false positive rates.
- Automatic seizure detection software is clinically viable but requires expert interpretation and awareness of software limitations.
More Related Videos
08:13SwarmSight: Real-time Tracking of Insect Antenna Movements and Proboscis Extension Reflex Using a Common Preparation and Conventional Hardware
Published on: December 25, 2017
06:04Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
Related Concept Videos
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Statgraphics
Comparison Tests