Cone beam computed tomography in dentistry: Clinical recommendations and indication-specific features
Related Experiment Video
Updated: Jun 16, 2025

Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
Rocharles Cavalcante Fontenele1, Hugo Gaêta-Araujo2, Reinhilde Jacobs3
1Department of Stomatology, Public Health and Forensic Dentistry, Division of Oral Radiology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, São Paulo, Brazil; OMFS-IMPATH Research Group, Department of Imaging and Pathology, Faculty of Medicine, KU Leuven, Leuven, Belgium; Department of Oral and Maxillofacial Surgery, UZ Hospitals, Leuven, Belgium.
Objectives:
To provide a narrative overview of commercially available CBCT devices, highlight the wide range of technical features and their impact on image quality and diagnostic performance, and develop decision support charts to guide clinicians in selecting appropriate CBCT parameters for different dental specialties.
Data And Sources:
Data collection involved reviewing published articles on CBCT devices and accessing manufacturer websites. Additional manufacturers were identified through a literature search and by attending scientific and commercial events held in 2023 and 2024. Information on CBCT features was obtained from websites, manuals, and inquiries until February of 2025. Data were categorized and analyzed descriptively and quantitatively to recommend device requirements for dental specialties. Decision support charts were developed for each specialty based on common indications and technical parameters (e.g., field-of-view, scout need, spatial resolution, X-ray exposure parameters, and post-processing tools).
Results:
Eighty-seven commercially available CBCT devices from 34 manufacturers across 11 countries were identified, with information confirmed for only 10 companies. Missing data varied, however FOV and voxel size were commonly reported. Tube voltage ranged from 50-120 kV, and tube current mostly varied between 1-17 mA. Dose considerations varied widely, with dose-product-area ranging from 10 to 5600 mGy.cm2 and typical effective dose estimates from 3 to 500 µSv. Decision support charts were created, outlining necessary image acquisition and reconstruction-related parameters for prosthodontics and implant dentistry, periodontology, oral and maxillofacial surgery, pediatric dentistry and orthodontics, and endodontics.
Conclusion:
This study provides an overview of the technical features of commercially available CBCT devices, highlighting gaps in reconstruction parameters and dose considerations. Tailored recommendations based on patient-specific needs are essential, with clinicians individualizing CBCT acquisition to optimize diagnostic accuracy and minimize biological risks.
Clinical Significance:
This study provides valuable insights into commercially available CBCT devices in terms of their features and provides decision support chart to help clinicians select optimal technical parameters tailored to specific dental specialties.
More Related Videos
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
09:10Digital Hybrid Model Preparation for Virtual Planning of Reconstructive Dentoalveolar Surgical Procedures
Published on: August 5, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...