Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Validity of Heart Rate Variability Measured with Apple Watch Series 6 Compared to Laboratory Measures
Lauren Bonneval1, David Wing2, Sydney Sharp2,3
1San Diego General Preventive Medicine Residency, Herbert Wertheim School of Public Health and Human Longevity Science, University of California San Diego, La Jolla, CA 92093, USA.
Abstract:
We assessed the test validity of the Apple Watch series 6 measure of heart rate variability (HRV) by comparing it with the reference measure assessed via a Biopac 3-lead electrocardiogram (ECG). We recruited 78 healthy adults (aged 20-75 years). HRV was measured using an in-lab protocol while resting, talking, watching a movie, before walking, and after walking. We conducted a synchronized countdown for each condition to guarantee that the recordings would be aligned between the two devices by using event markers in the Biopac at the exact time that the Apple Watch Breathe app began and ended. We assessed test validity using the Bland-Altman method, and both precision and accuracy were estimated using Lin's concordance correlation coefficient. The highest level of agreement and concordance between devices occurred during rest. We observed near-perfect agreement for R-R intervals and beats per minute (BPM) measures, with mean absolute percentage errors (MAPE) of 1.15% during resting conditions. We observed moderate levels of agreement and concordance for N-N intervals at rest with a MAPE of 31.31% during resting conditions. The Apple Watch provides a high level of validity for measuring R-R intervals and BPM in healthy adults. Further research is needed to determine if HRV measures with the Apple Watch offer a significant opportunity for the surveillance of CVD risk.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)