A Novel Column-Switching Method Coupled with Supercritical Fluid Chromatography for Online Analysis of Bisphenol A
Chaoyan Lou1,2, Shaojie Pan1, Kaidi Zhang1
1College of Quality and Standardization, China Jiliang University, Hangzhou 310018, China.
Abstract:
Bisphenol A diglycidyl ether (BADGE) and its related derivatives (BADGEs for short) are reactive epoxides condensed from bisphenol A (BPA) and epichlorohydrin. Nowadays, they are heavily used as additives in the production process of food and beverage contacting materials. However, BADGEs are considered as emerging organic pollutants due to their high toxicity including cytotoxicity, mutagenicity, and genotoxicity. In this work, an online analytical method integrated column-switching technique with supercritical fluid chromatography (SFC) was proposed for the simultaneous determination of bisphenol A diglycidyl ether and its derivatives. In this process, a homemade column was utilized in the first dimension of the column-switching SFC system to preconcentrate the analytes as well as eliminate interferences online. Under the optimal conditions, the obtained calibration curves for BADGEs showed good linearity ranging from 0.02 μg/mL to 10.00 μg/mL, while the values of LOD and LOQ were in the range of 0.0024-0.0035 μg/mL and 0.0080-0.0116 μg/mL, respectively. The optimized method exhibited a good recovery ranging from 85.6% to 105.5% with relative standard deviations less than 11.8%. The developed method provides an eco-friendly and effective way for the rapid and automated analysis of BADGEs at trace levels in canned beverages and can be applied to the high-throughput analysis of other similar matrices.
More Related Videos
Related Concept Videos
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
High-Performance Liquid Chromatography: Instrumentation