Radiation dose reduction and image quality enhancement for patients unable to elevate their arms in chest CT: A
Ki Baek Lee1, Seong Min Kim2, Jong-Tae Yoon3
1Department of Radiologic Technology, Chungbuk Health & Science University, Cheongju 28150, Republic of Korea.
Purpose:
To propose a method to effectively reduce radiation dose while enhancing image quality in chest CT for patients unable to elevate their arms utilizing the air-gap technique.
Methods:
Chest CT images were acquired in three positions: arms raised (control group, position A), arms lowered with the air-gap technique utilizing an in-house device (position B), and arms lowered without the device (position C). Data were categorized by body weight. Radiation dose was assessed using dose-length product (DLP) values, with quantitative analyses of image noise, contrast-to-noise ratio (CNR), and CNR-to-dose ratio (CNRDR). Qualitative assessment utilized a Likert scale.
Results:
Average DLP was 473.7 ± 142.9 mGy × cm in position C, with the lowest in position A at 267.3 ± 93.3 mGy × cm. Position B had a DLP of 317.2 ± 121.4 mGy × cm, 18.7 % higher than position A but 33.0 % lower than position C. In patients over 70 kg, all positions exceeded the thoracic CT diagnostic reference level (DRL) (324.2 mGy × cm). For those under 70 kg (i.e., <50 kg, 50-70 kg), position B showed lower DLPs compared to the above DRL. Quantitative analysis indicated that position B maintained image quality similar to position A, with CNRDR of 30.2 ± 11.1 versus 26.6 ± 10.9 in position A, showing marginal significance (p = 0.046). The qualitative evaluation indicated that position A had the highest rating at 4.96, followed by position B at 4.51, and position C at 3.47.
Conclusion:
The air-gap device with lowered arms during chest CT for patients unable to elevate their arms can reduce radiation dose while maintaining image quality. A custom-designed device facilitates safe and efficient procedures, especially for patients with higher body weight.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging