An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Comparison of fracture resistance of implant-supported fixed prothesis substructure materials with different
Oğuzhan Yılmaz1, Ayşegül Göze Saygın2, Giray Bolayır1
1Faculty of Dentistry - Department of Prosthodontics, Sivas Cumhuriyet University, Sivas, Turkey.
Background:
The aim of this study was to compare fracture resistance after thermomechanical ageing of prosthetic substructure materials with different connector designs.
Methods:
Three different prosthetic substructure materials were used in this study: (1) GroupZir: (Zirconia, Fusion ceram, Turkey), (2)GroupPEEK (PEEK, Whitepeaks, CopraPeek, Essen, Germany), (3)GroupFRC (Fibre-reinforced composite, Trinia, Bicon Implant, Rep. of Ireland). A total of 72 implant-supported prosthesis triangular, square, and oval connector designs were created between 2nd premolar and mandibular 2nd molar teeth. After adhesion to the implant abutments with resin cement (Pentron breeze, Kerr), the samples were applied with dynamic loading and thermomechanical ageing (120,000 cycles,120 N,5-55 °C). Fracture resistance values were obtained with a universal test device and SEM images were analysed. The analyses were performed with Two-Way ANOVA and the Tukey test (SPSS 23.00).
Results:
Both the material and the connector type were found to affect the fracture resistance (F = 8.354, p < 0.05). The highest fracture resistance value was obtained from the triangular shape in GroupZir(3200 ± 91.05) and the lowest from the oval connector design of GroupPEEK material (2410 ± 157.23). Statistically significant differences were determined in the different connector designs of GroupZir(p < 0.05). In the comparisons made according to connector design, a significant difference was obtained between GroupZir and GroupPEEK and between GroupZir and GroupFRC. Deformations were observed in the fracture pattern of the Group Zir samples and deformations in the form of rupture were seen in the GroupPEEK and GroupFRC material samples.
Conclusions:
The study results demonstrated that the fracture resistance of zirconia, PEEK, and FRC restorations over 3-unit implants with different connectors is affected by connector design. All the materials were seen to be comparable in respect of the forces formed in chewing dynamics.
More Related Videos
09:08Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Stress-Strain Diagram - Brittle Materials
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...
Plastic Deformations
Behavior of Concrete Under Compressive Load
As the concrete specimen fractures under...
Types of Non-structural Cracks in Concrete
Plastic shrinkage cracks typically form within hours after the concrete is poured. The concrete's surface dries faster than the bottom, creating tensile stress that the still-plastic concrete cannot withstand, leading to diagonal or randomly patterned cracks on the concrete surface.
Plastic...
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used to...