[National Survey on System Settings of Angiography for Endovascular Treatment for Lower Extremity Arterial Disease]
Yuji Sakai1, Tomomichi Ishibashi2, Yoshinori Takao1
1Department of Radiology, Osaka Metropolitan University Hospital.
Purpose:
The objective of this study was to determine the standard settings of the angiography system during EVT in Japan.
Methods:
Survey requests were mailed to 493 facilities across Japan. The survey items included in the survey were system display dose setting of fluoroscopy dose (FL dose)/digital angiography dose (DA dose)/digital subtraction angiography dose (DSA dose), detector size, fluoroscopy pulse rate, DA/DSA frame rate, and standard additional filter settings. The number of valid responses for each item was system display dose setting: 187 systems (128 institutions), detector size: 187 systems (128 institutions), fluoroscopy pulse rate and DA/DSA frame rate: 185 systems (127 institutions), standard additional filter settings: 182 systems (125 institutions).
Results:
Median (interquartile range) of each system display dose setting was FL: 5.9 (4.2-7.8) mGy/min, DA: 0.10 (0.06-0.16) mGy/frame, DSA: 0.82 (0.64-1.40) mGy/frame. The detector size for dose measurements was 103 systems (55%) with a diagonal of 35 cm or less and 84 systems (45%) with a diagonal of more than 35 cm. The most used FL pulse rate was 7.5 pulse/s (132 systems: 71%), DA frame rate was 7.5 frame/s (72 systems: 39%), and DSA frame rate was 3 frame/s (97 systems: 52%). More than 90% of the institutions used some kind of additional filter for imaging.
Conclusion:
This survey showed the standard setting of the angiography system during EVT in Japan.
More Related Videos
14:52Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
Published on: December 11, 2013
07:25Predicting Amputation using Local Circulating Mononuclear Progenitor Cells in Angioplasty-treated Patients with Critical Limb Ischemia
Published on: September 22, 2020
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...