Hypotension prediction index: comparison between invasive and non-invasive pressure inputs
Kamron Sarhadi1, Justin Hamman2, Jorge Avila3
1Department of Anesthesia & Perioperative Care, University of California, San Francisco, San Francisco, CA, USA.
Background:
The Hypotension Prediction Index (HPI) derived using an Acumen™ arterial pressure transducer can decrease the incidence of intraoperative hypotension and possibly decrease perioperative complications. The HPI can also be obtained from the ClearSight™ continuous non-invasive blood pressure monitor. Concurrent comparison of HPI values obtained from these two pressure inputs is limited and additional comparisons could increase clinician confidence in non-invasive hemodynamic monitoring, expand its applications and improve patient outcomes.
Methods:
Simultaneous hemodynamics were recorded using two HemoSphere monitors with the HPI software and either intra-arterial Acumen® (hereinafter invasive) or ClearSight™ (hereinafter non-invasive) pressure inputs. Data collected from the non-invasive system was compared to corresponding invasive, intra-arterial data using Bland-Altman analysis, Spearman correlation, concordance analysis and relative performance with respect to prediction of hypotensive events using ROC analysis and HPI alerts agreement analysis.
Results:
6,862 paired data points were available from 36 patients. Bland-Altman comparisons demonstrated a bias of -8.4 (± 23) with limits of agreement from - 53 to 36. The correlation between HPI values was strong with an r value of 0.76 (95%CI:0.75-0.77). Concordance was also strong at 74% (10% exclusion zone). Using ROC analysis, the AUC for prediction of hypotension was similar and at 5 min was 0.883 [0.786,0.953] for the invasive pressure and 0.860 [0.770,0.939] for the non-invasive pressure inputs. At the same time points, the agreement between HPI alerts was high with an accuracy of 86.3%.
Conclusion:
HPI values and predictive performance were comparable when derived from either invasive or non-invasive pressure inputs.
Trial Registration:
The study was approved by the UC Davis Institutional Review Board (IRB #1791102-1) and registered on ClinicalTrials.gov (NCT05025176) before the enrollment of the first patient.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure