Related Experiment Video
Updated: May 9, 2025

Evaluation of Biomaterials for Bladder Augmentation using Cystometric Analyses in Various Rodent Models
Published on: August 9, 2012
Ensuring patient safety during cystocopy: risk assessment of device reprocessing through healthcare failure mode and
Miriam Roncal Redin1,2,3, Cristina Diaz-Agero Perez4,5,6, Cornelia Bischofberger Valdes4,5
1Hospital Universitario Ramón y Cajal, Ctra. de Colmenar Viejo, km. 9,100, 28034, Madrid, Spain. miriam.roncal@salud.madrid.org.
Introduction:
Flexible cystoscopy is a widely performed urological procedure, with millions conducted annually. However, cystoscope contamination rates are relatively high, ranging from 2 to 21%, and infections may go undetected. Proper reprocessing of cystoscopes is a critical step in ensuring the safe use of these devices, which underscores the importance of studying it more thoroughly. The aim of this article is to describe the stages of cystoscope reprocessing in the Urology Department of a tertiary Hospital, identifying the potential failures of the system before they occur, quantifying its risks and recommending improvement actions to enhance overall safety and efficiency.
Material And Methods:
This qualitative study was conducted using the Failure Mode and Effects Analysis methodology in different phases. First, data on structure, equipment and work processes were collected through direct observation and staff interviews. Then, processes and sub-processes were described along with their possible failures and underlying causes. Risks of this potential failures (failure modes) were quantified using the Risk Priority Number (RPN), calculated as the product of their frequency, severity and detectability. Improvement actions to mitigate failures with the highest NRP were proposed based on expert consensus and literature review.
Key Findings And Limitations:
9 processes and 11 sub-processes were identified, along with 16 failure modes. The key failure modes, those with an RPN > 100 included incorrect manual high-level disinfection, incorrect cleaning due to lack of brushing and lack of traceability. These failures were primarily caused by insufficient staff training, short intervals between procedures, and inadequate infrastructure. To address these issues, expert recommendations included providing comprehensive training for staff, improving reprocessing infrastructure by optimizing room layout, and ensuring an adequate supply of cystoscopes to reduce time constraints. Additionally, prioritizing automatic high-level disinfection and implementing a traceability system were proposed to enhance process reliability and patient safety.
Conclusion:
Inadequate organizational structure and equipment can compromise patient safety during high-care activities. After analyzing cystoscope reprocessing at our center, three critical failure modes (NPR > 100) were identified: lack of traceability, improper cleaning due to insufficient brushing, and improper manual high-level disinfection (HLD). To address these issues, we proposed actions including improve staff training, increase the cystoscope supply, and enhance the unit's infrastructure. These findings may guide other healthcare professionals in improving reprocessing safety.
Related Concept Videos
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Asepsis
Healthcare Associated Infections I: Iatrogenic, Exogenic and Endogenic
HAIs significantly increase the cost of health care. Extended stays in healthcare institutions, increased disability, increased costs of medications, including specialized antibiotics, and prolonged recovery times add to the patient's expenses and the healthcare institution and funding bodies.