Erroneous calibration of esophageal pressure in case of airway closure
Mattia Docci1,2, Francois Beloncle3, Arnaud Lesimple3,4
1Keenan Centre for Biomedical Research, Li Ka Shing Knowledge Institute, Unity Health Toronto, Toronto, ON, Canada.
Abstract:
Airway closure results in a lack of communication between proximal and distal airways unless the airway pressure (Paw) overcomes the airway opening pressure (AOP). This has been described in patients undergoing mechanical ventilation with acute respiratory distress syndrome, obesity, hydrostatic pulmonary edema and during cardiopulmonary resuscitation. In these categories of patients, esophageal pressure (Pes) can guide the personalization of mechanical ventilation and calibration of the esophageal balloon is necessary to obtain reliable Pes measurements. The impact of airway closure has never been envisaged. This study investigated the impact of airway closure on the calibration of the esophageal balloon by the ∆Paw/∆Pes following a positive pressure occlusion test during passive mechanical ventilation. The calibration test was performed in twelve human cadavers with airway closure at end-expiration at different levels of positive end-expiratory pressure (PEEP) and at end-inspiration. The ∆Paw/∆Pes measured at end-expiration and at end-inspiration were significantly different when total PEEP was lower than AOP (estimated means 0.42 [0.40; 0.44] vs. 0.95 [0.92; 0.97], P < 0.001), while this difference was not observed when total PEEP was higher than AOP (estimated means 0.99 [0.92; 1.05] vs. 0.99 [0.92; 1.06], P = 0.854). These results were corroborated by observations during esophageal balloon calibration in two patients requiring Pes monitoring for clinical management. In case of airway closure, compression of the chest is not fully transmitted to the airways. This can lead to a conspicuous underestimation of the ∆Paw/∆Pes and poor reliability of this monitoring technique when the test takes place below AOP. Our results favor a positive pressure occlusion test performed during an end-inspiratory occlusion as the new standard of operative procedures for positioning and calibrating the esophageal balloon.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Esophageal Strictures-I: Introduction
Etiology
The primary cause of esophageal strictures is long-standing gastroesophageal reflux disease (GERD), accounting for about 70 to 80% of adult cases. Chronic acid reflux can lead to injury and scarring of the esophageal lining, culminating in...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Measurement of Blood Pressure
Esophageal Perforation-I: Introduction
The location of esophageal perforation can vary, occurring anywhere along the esophagus....