A Wider Grip: Endoanchors Increase Proximal Sealing Surface during Endovascular Aneurysm Repair
Giulio Accarino1, Davide Esposito2, Marco Campolmi3
1Vascular Surgery Unit, Struttura Ospedaliera ad Alta Specialità Mediterranea, Naples, Italy; Department of Public Health, Vascular Surgery Unit, University Federico II of Naples, Naples, Italy; Department of Medicine, Surgery and Dentistry, University of Salerno, Baronissi, Salerno, Italy.
Background:
To evaluate how the Heli-FX EndoAnchor system (Medtronic Vascular, Santa Rosa, CA, USA) influences the Real Achieved Sealing Zone (RASZ) compared to Endovascular Aneurysm Repair (EVAR) with Endurant (Medtronic Vascular, Santa Rosa, CA, USA) alone in a matched and unmatched real-world cohort.
Methods:
In a retrospective, single-center study (2015-2022), 221 Abdominal Aortic Aneurysm patients underwent Endurant stent graft (Medtronic Vascular, Santa Rosa, CA, USA) placement either alone (n = 190) or with adjunctive Heli-FX EndoAnchors (n = 31). Preoperative and postoperative computed tomography angiography scans were evaluated using a standardized measurement technique (closed-polygon area, center lumen line distances) to quantify the Target Anticipated Sealing Zone (TASZ) and the RASZ. Primary endpoints were changes in RASZ length and area compared to TASZ. Secondary endpoints included type 1A and any endoleak rates, overall reinterventions. Statistical analysis included unadjusted and adjusted Cox proportional hazard regression models, receiver operating characteristic (ROC) curves, and propensity score matching.
Results:
Heli-FX patients presented significantly more "hostile" proximal sealing zones, featuring shorter TASZ length (11 vs. 17 mm; P = 0.003), higher infrarenal angulation (45° vs. 18°; P < 0.001). Despite this hostile anatomy, the Heli-FX group exhibited a substantial increase in RASZ area (+48.3 mm2; P = 0.001) and length (+5.8 mm; P = 0.003) compared to standard EVAR patients. ROC analysis indicated distinct RASZ cutoffs for predicting type 1A endoleak in each group (27.7 mm2 and 4.5 mm for Heli-FX vs. 49.5 mm2 and 9.5 mm for Endurant). In the 24 TASZ-features-matched couples, Heli-FX patients showed increased RASZ area (+42 vs -31 mm2 over TASZ; P = 0.003), and length (+5 vs -3 mm over TASZ; P = 0.003).
Conclusion:
EndoAnchors substantially improve RASZ length and surface area compared to EVAR with the same stent graft alone. Routine device-specific RASZ assessment could refine treatment algorithms and pave the way for more personalized and durable endovascular solutions.
More Related Videos
Related Concept Videos
Anchoring Junctions
Regulation of Angiogenesis and Blood Supply
Desmosomes
Tight Junctions
Adherens Junctions
Adherens Junctions are Dynamic
GPI Anchoring of Proteins in the ER Membrane
GPI-anchor structure
A sequence of 11 enzymatic reactions results in the synthesis of the complete GPI anchor consisting of a hydrophobic and a hydrophilic portion. The hydrophobic portion comprises phosphatidylinositol, while the hydrophilic part comprises polar groups like phosphoethanolamine,...