A new design to improve time resolution in a time of flight brain PET using dual layer offset scintillator crystals
Amir Beheshti1, Alireza Karimian2, Hossein Arabi3
1Faculty of Physics, University of Isfahan, Isfahan, Iran. beheshti.amir86@gmail.com.
Abstract:
In PET systems, the SNR relies on the coincidence time resolution (CTR) of 511 keV photon pairs. This research investigates the impact of reflectors, surface treatments, materials, and scintillation crystal length on the CTR of a brainPET detector using dual-layer offset scintillators (DLOs). This study is based on a brainPET, under development at the University of Manitoba, to propose a new design to achieve an improved CTR. Four different pairs of LYSO crystals with distinct optical compositions, surface treatments, and reflective materials were simulated (using GATEv9.3). Each model comprises two LYSO crystal with dimensions of 3 × 3 × 10 mm3. Considering the initial experimental data from the brainPET lab, simulation results showed that the crystal with a roughened surface and ESR reflector demonstrated 13.6% energy resolution and an average 17.8% improvement in CTR compared to other models. In addition, a more comprehensive model, including a dual-layer offset detector was designed. The bottom and top layers have 25 × 19 and 24 × 18 crystals with thickness of 12 and 8 mm, respectively in the DLO model. The simulation investigation showed that the DLO configuration could enhance the time resolution by 17.5% and the energy resolution by 5.4% which are considerably comparable to the state-of-the-art brainPET systems.
More Related Videos
09:07Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016