Validation assessment of a cuff-based central blood pressure device according to Artery Society recommendations
Tan V Bui1, Dean S Picone1,2, Wojciech P Kosmala1,3
1Menzies Institute for Medical Research, University of Tasmania, Hobart.
Objective:
Uscom BP+ is a cuff-based blood pressure (BP) device designed to noninvasively estimate central BP as distinct from conventional brachial BP. This study aimed to assess the accuracy of the Uscom BP+ device compared with invasively measured BP.
Methods:
Automated noninvasive cuff central BP (using the Uscom BP+ device) and invasive central aortic BP were recorded simultaneously in 191 participants (65% male, aged 66 ± 11 years) receiving coronary angiography at three independent research sites in Australia, Poland, and Italy. Validation procedures were undertaken according to the Artery Society recommendations and with a minimally acceptable error (mean ± SD) of ≤5 ± ≤8 mmHg as pass criteria.
Results:
Using the device default calibration technique [brachial cuff systolic blood pressure (SBP) and diastolic blood pressure (DBP)], cuff central SBP underestimated invasive central SBP [mean (SD) difference: -10.2 (11.2) mmHg] and cuff central DBP overestimated invasive central DBP [mean (SD) difference: 9.8 (8.5) mmHg]. When calibrating by brachial cuff mean arterial pressure and DBP, SBP accuracy was improved, but variability remained high [mean (SD) difference: -6.3 (14.4) mmHg, P = 0.004 vs. default calibration, whereas DBP accuracy and variability remained similar [mean (SD) difference: 10.9 (8.5) mmHg, P = 0.19 vs. default calibration].
Conclusion:
The Uscom BP+ cuff device does not pass the Artery Society accuracy criteria compared with invasively measured central BP.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...