The "16-gram window" of contact-force: A new criterion for very high-power short-duration ablation
Dimitrios Bismpos1,2, Jan Wintrich1,2, Valerie Pavlicek1
1Department of Internal Medicine III, Cardiology, Angiology and Intensive Care Medicine, University Hospital Saarland University Homburg Saar Germany.
Background:
Very high-power short-duration (vHPSD) ablation with the novel QDOT™ catheter allows the regulation of target temperature by automatically adjusting flow and power during a 4 s application of 90 W. However, the optimal contact force for sufficient lesion creation is unknown.
Methods:
We enrolled 73 patients with symptomatic atrial fibrillation undergoing pulmonary vein isolation (PVI) using the QDOT catheter in the vHPSD mode (90 W, 4 s). Ablation metrics associated with suboptimal applications, defined as either an impedance drop of ≤5% or a cumulative temperature-limited energy ≤330 J, were collected and analyzed.
Results:
A total of 3881 vHPSD applications (53.2 applications per patient) with a mean contact force (CF) of 12.8 ± 6.6 g were analyzed. Significant CF variability and intermittent loss of contact were documented in 18.2% and 8.8% of the applications, respectively. A ΔImp ≤ 5% occurred in 3.9% of vHPSD applications, while a cumulative energy ≤ 330 J was observed in 3% of the applications. Applications with a mean CF < 6 g and >22 g were associated with an inadequate impedance drop (10.3%, Phi coefficient 0.118, p < .001) and total applied energy (7.8%, Phi coefficient 0.094, p < .001) respectively. At superior PV segments with thick atrial walls, significantly more applications with cumulative energy ≤330 J (4.2% vs. 2.5%; p = .007) were observed, especially when mean CF > 18 g was applied (8.4%, Phi coefficient 0.093, p = .003).
Conclusion:
A lower but also a higher mean contact-force was associated with suboptimal vHPSD applications. Hence, a "16-gram window" of contact-force, from 6 to 22 g, could optimize energy application in vHPSD ablation.