Accuracy of Gravity-Based Automatic Infusion System for Chemoport Intravenous Infusion
Young Gyu Park1, Jong-Yeup Kim2, Dong Won Yang2
1Department of Oncology-Hematology, Konyang University Hospital, Daejeon, Republic of Korea.
Background:
An infusion pump designed to deliver a specific volume at a predetermined rate is unable to detect extravasation and may continue infusion despite increased tissue resistance caused by extravasation. Consequently, it is recommended that vesicant drugs, such as chemotherapeutic agents, be administered via gravity infusion to reduce the risk of tissue damage. However, intravenous flow regulators used for gravity infusion have limitations because the infusion rate is influenced by the height, venous pressure, and viscosity of the fluid, which can change with temperature. In this study, we measured the accuracy of Accudrip, a new gravity-based automatic infusion control device designed to correct fluctuating infusion rates.
Methods:
In 59 cancer patients administered with anticancer drugs, the actual administration rate using Accudrip under various administration conditions and the theoretical administration rate were measured and analyzed 100 times.
Results:
Comparing the theoretical and actual administration rates using Accudrip confirmed that it can be injected with an average error rate of 4.75%, which clinically demonstrated sufficient accuracy for fluid administration.
Conclusion:
The gravity-based automatic infusion system showed high accuracy consistent with the theoretical rate of administration even under multiple infusion conditions of cancer patients with chemoport. This system (Accudrip) will greatly contribute to accurate drug administration and minimization of adverse events in real-world hospital settings.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Compartment Models: Single-Compartment Model
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...