Usefulness of high tube voltage conditions in CT Fluoroscopy during CT-guided biopsy: Preliminary study
Kento Furihata1, Jun Miyagawa1, Naomichi Tamaru1
1Division of Radiology, Shinshu University Hospital, Matsumoto, Nagano, Japan.
Objectives:
Computed tomography (CT)-guided biopsy is often used to increase the safety and accuracy of biopsies for disease diagnosis. However, CT-guided biopsy is associated with metal artifacts from the biopsy needle and increased patient and operator exposure due to frequent CT fluoroscopy. Therefore, we thought it possible to solve this problem by setting the CT fluoroscopy conditions to a higher tube voltage and a lower tube current-time product (high-tube voltage conditions) than recommended. As a preliminary study, metal artifacts, low-contrast detectability, patient and operator's exposure, and visual changes in high-tube voltage conditions were assessed using phantoms and compared with recommended conditions.
Material And Methods:
On an interventional radiology CT system, the phantom was scanned under recommended conditions (120 kV, 30 mAs) and high-tube voltage conditions (135 kV, 30-5 mAs). The metal artifacts and low-contrast detectability of each condition were analyzed and compared using the acquired images. In addition, the phantom surface dose assuming patient exposure and the air dose assuming the operator's standing position were measured and compared. Furthermore, visual assessment was performed by six radiological technologists.
Results:
Low-contrast detectability was slightly reduced, metal artifacts were significantly lower under high-tube voltage conditions, and patient and operator exposure were lower than the recommended conditions. Furthermore, the findings of the visual assessment were largely consistent with those of the physical assessment.
Conclusion:
High-tube voltage conditions in CT fluoroscopy during CT-guided biopsy may be useful in reducing metallic artifacts and patient and operator radiation exposure.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...