High-Performance Liquid Chromatography Using Ultraviolet Detection for Separation of Terephthalic Acid and Associated
Mehdi Ashraf-Khorassani1, William M Coleman Iii2, Tripp Aardema1
1Department of Chemistry, 1040 Drillfield Drive, Virginia Tech, Blacksburg, VA 24061.
Abstract:
An optimized rapid and precise high-performance liquid chromatography method has been developed to separate, identify and quantitate impurities contained within terephthalic acid (TPA) produced from recycled materials. Using a Waters X-Select HSS T3 column for the separation of recycled TPA and eight impurities, and a mobile phase of H2O/CH3CN containing 0.1% trifluoroacetic acid (TFA) with detection by UV at 250 nm, baseline separation of all compounds was achieved in ˂17 min. Four different mobile phase acidic additives were tested, the results indicating that mobile phases with lower pH provided better separation of the compounds with accompanying improved peak shapes. Among all the acidic additives tested in this study, a mobile phase containing 0.1% TFA demonstrated the highest resolution and the best peak shapes. It was concluded that by suppressing the ionization of the acidic compounds contained in the samples through increased acidity of the mobile phase, better peak shapes and higher resolution could be achieved. The amount of increased mobile phase acidity was based in part on the pKa values of the analytes and adjusted so as to eliminate as much as possible any dissociation of the organic acids. A comparison of a standard addition method versus an external calibration curve approach for the quantification of impurities showed similar results, even for relatively low-concentration analytes at levels below 10 ppm. The aspects often linked with validated analytical methods were addressed including linearity, limit of detection, limit of quantitation, sensitivity and selectivity, addressing the criteria set forth in the acronyms of the QuECHERS method description, that is quick easy, cheap, effective, rugged and safe.
More Related Videos
13:02Arabidopsis thaliana Polar Glycerolipid Profiling by Thin Layer Chromatography TLC Coupled with Gas-Liquid Chromatography GLC
Published on: March 18, 2011
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Elution Process
Thin-Layer Chromatography (TLC): Overview
To begin the analysis, a mixture of compounds is spotted on the starting line on the TLC plate using a thin capillary. The bottom of the...