Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Small-field output factor dependence on the field size definition in MR-Linac
Indra J Das1, Ahtesham U Khan1, Poonam Yadav1
1Department of Radiation Oncology, Northwestern Memorial Hospital, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA.
Background:
Radiation beam characteristics are primarily evaluated based on field size. However, in small fields, especially with magnetic fields used in new technology (MR-Linac), the field size definition is altered. Typically, field size is defined by two methods: geometric and dosimetric, which are evaluated in this study.
Purpose:
Small field size definitions are distorted due to lateral electron disequilibrium and the presence of magnetic fields. MR-Linac systems, which combine an MR imaging system and a linear accelerator on a single gantry, require precise evaluations of field size definitions and beam parameters, particularly for small fields. which is investigated in this study.
Methods:
A 0.35 T MRIdian Viewray system was evaluated using beam profiles and field output factors (FOF) with various MR-compatible microdetectors, such as ion chamber, microDiamond, microSilicon, and plastic scintillators. Validity of geometric field size (S) and dosimetric field size (Sclin) is investigated with measurements performed with MR compatible scanning water phantom at 85 cm source-to-surface distance (SSD) at a depth of 5 cm. Measured FOF data was compared with treatment planning systems (TPS) and independent Monte Carlo simulations.
Results:
The measured Sclin data is detector and machine-dependent, while S is machine-dependent only. The FOF was found to be a smooth function of S within experimental uncertainties, showing higher reproducibility compared to Sclin which exhibited erratic behavior.
Conclusions:
It is concluded that geometric field size (S) provides accurate beam characterization data, whereas Sclin may not be a reliable parameter in MR-Linac systems.
Related Concept Videos
Induced Electric Fields: Applications
NMR Spectrometers: Resolution and Error Correction
Electric Field of a Continuous Line Charge
In calculations of electric fields, symmetry is of great use. For example, while calculating electric fields of continuous charge distributions.
Consider a line element with a...
Induced Electric Fields
Faraday's Law
Motional Emf