Optimization of MALDI Matrices and Their Preparation for the MALDI-TOF MS Analysis of Oligonucleotides
Susanne Dietrich1,2, Anja Dollinger1,3, Andreas Wieser2,3,4,5
1Chair of Analytical Chemistry, TUM School of Natural Sciences, Technical University of Munich, Munich, Germany.
Rationale:
The reproducibility of the analysis of oligonucleotides using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) remains a significant challenge. This is mainly attributed to factors such as the choice and application of the matrix, as well as the inhomogeneity of the sample spots. Although previous studies have explored various parameters, such as ionic and sugar-based matrices, and different spotting methods, none have comprehensively integrated these factors into a unified approach.
Methods:
Various matrices for the analysis of oligonucleotides and the effects of diverse variables, including different matrices, solvent compositions, additives, and application techniques, on the analytical performance of these matrices were investigated. Mass spectrometry analysis was conducted by a MALDI-TOF MS in linear negative mode.
Results:
Our research systematically evaluates the combined effects of diverse variables to enhance the analytical performance of MALDI-TOF MS in oligonucleotide analysis. We focused on the standard deviation of mass-to-charge ratios and the signal-to-noise (S/N) ratios. Out of 48 samples, only 19 met the S/N criteria, which is that the signals must be detectable over the whole mass range of interest (4-10 kDa).
Conclusions:
The ionic matrix 6-aza-2-thiothymine (ATT) with 1-methylimidazole resulted consistently in a reduced standard deviation and achieved high mass precisions. Additionally, we observed that the S/N ratios and mass precision of 3-hydroxypicolinic acid (3-HPA) varied significantly depending on the solvent composition and the presence of additives.
More Related Videos
07:49Matrix-assisted Laser Desorption/Ionization Time of Flight MALDI-TOF Mass Spectrometric Analysis of Intact Proteins Larger than 100 kDa
Published on: September 9, 2013
06:56Characterization of Synthetic Polymers via Matrix Assisted Laser Desorption Ionization Time of Flight MALDI-TOF Mass Spectrometry
Published on: June 10, 2018
Related Concept Videos
Matrix-Assisted Laser Desorption Ionization (MALDI)
The analyte of interest, a biomolecule or a mixture of biomolecules, is mixed with a suitable matrix material. The...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...