Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Validation of portable ultrasonic stadiometers in adults
Joaquim Huaina Cintra-Andrade1, João Victor Alves Martins2, Maria Izabel Freire-Correia3
1Federal University of Ceará, Fortaleza, Ceará, Brazil.
Background:
Over the past two decades, alternatives to measuring standing stature using portable devices based on innovative technologies have been explored. This study aimed to assess the validity of three portable ultrasonic stadiometers for measuring unstretched standing stature in comparison to a portable floor-standing mechanical stadiometer in Brazilian adults.
Methods:
Cross-sectional study carried out with a convenience sample of 115 individuals (51.3 % women) with a mean age of 31.5 ± 9.3 years and able to stand independently. The portable ultrasonic stadiometers Avanutri®, Balmak® and Inbody® were compared to the portable mechanical floor-standing stadiometer Avanutri®, which was treated as a reference, to measure unstretched standing stature. Differences between mechanical and ultrasonic stadiometers were analyzed by paired-sample t-tests with 95 % confidence intervals (95 % CI). Furthermore, the equivalence level was assessed using the Two One-Sided Test (TOST) procedure.
Results:
Statistically equivalent absolute mean differences were observed in the comparative analysis between the stadiometers. The Avanutri® and Balmak® ultrasonic stadiometers produced significant systematic bias of 0.79 cm (95%CI: 0.471; 1.116) and 0.40 cm (95%CI: 0.084; 0.715), respectively. The Inbody® ultrasonic stadiometer produced a systematic bias of -0.03 cm (95%CI: -0.291; 0.229), below the 1 mm limit.
Conclusion:
The Inbody® portable ultrasonic stadiometer demonstrated validity for accurately measuring standing stature in Brazilian adults.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement