The Reliability of a Novel Structured Testing Device for Single-breath Lung Carbon Monoxide Uptake: A Randomized
Shihua Yao1, Tuping Fu2, Kuiqing Lin1
1National Clinical Research Center for Respiratory Disease, State Key Laboratory of Respiratory Disease, Guangzhou Institute of Respiratory Health, The First Affiliated Hospital of Guangzhou Medical University, 151 Yanjiang Road, Guangzhou, China.
The MeHow MeAir 9000 spirometer accurately measures lung diffusing capacity and reliably assesses impairment severity, proving suitable for clinical use in patients with ILD and COPD.
Area of Science:
- Pulmonary Function Testing
- Respiratory Diagnostics
- Medical Device Validation
Background:
- New domestic lung diffusing capacity testing instruments are emerging.
- Clinical validation of the MeHow MeAir 9000 spirometer is necessary.
- The MeAir 9000 features a novel turning valve structure.
Purpose of the Study:
- To evaluate the clinical reliability of the MeHow MeAir 9000 spirometer.
- To assess the accuracy of lung diffusing capacity measurements.
- To compare MeAir 9000 performance against a established device.
Main Methods:
- 166 participants (30 healthy, 68 ILD, 68 COPD) were enrolled.
- A crossover design compared MeAir 9000 with Jaeger MasterScreen Diffusion.
- Lung diffusing capacity for carbon monoxide (DLCO) and DLCO %pred were primary outcomes.
Main Results:
- High consistency was observed for DLCO (CCC=0.988) and DLCO %pred classification (ICCs 0.931–0.975).
- Secondary indicators showed fair to poor consistency.
- MeAir 9000 demonstrated better CH4 detection stability, while MasterScreen showed superior CO detection.
Conclusions:
- The MeHow MeAir 9000 spirometer offers high accuracy in lung diffusing capacity measurements.
- It reliably assesses the severity of diffusing capacity impairment.
- The device is suitable for clinical application in respiratory diagnostics.