Comparison of image quality evaluation methods for magnetic resonance imaging using compressed sensing-sensitivity
Norikazu Koori1, Shohei Yamamoto2, Hiroki Kamekawa3
1Department of Radiological Technology, Faculty of Medical Technology, Niigata University of Health and Welfare, 1398 Shimami-cho, Niigata city, Niigata, 950-3198, Japan. Norikazu-koori@nuhw.ac.jp.
Abstract:
This study aimed to compare the relationship between the quantitative values and visual score of acquired images using the CS-SENSE method. T1-weighted image (T1WI) and T2-weighted image (T2WI) were acquired using a phantom created by a 3D printer. Each quantitative values (signal-to-noise ratio [SNR], contrast-to-noise ratio [CNR], structural similarity [SSIM], and scale-invariant feature transform [SIFT]) and visual evaluation score (VES) were calculated by the acquired images. The correlation coefficients among the calculating quantitative values and VES were calculated. The difference in methods for evaluating the image quality of T1WI and T2WI images using CS-SENSE was clarified. Variations in image quality, as reflected by VES in T1WI and T2WI images obtained via the CS-SENSE method, can be quantitatively assessed. Specifically, CNR is effective for evaluating changes in T1WI, while SNR, CNR, and SIFT are suitable for assessing variations in T2WI.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...