Test-retest repeatability of quantitative organ and tissue uptake using 20-minute dynamic multiparametric whole-body
Jonathan M Baier1,2, Kristian L Funck2,3, Anna Dons-Jensen2
1Dept. of Clinical Medicine, Aarhus University, Aarhus, Denmark.
Background:
Recently developed dynamic whole-body PET/CT (D-WB PET/CT) protocols allow for measurements of potentially more precise metabolic parameters than the commonly used semiquantitative SUV. Most notable is the metabolic rate of FDG uptake (MRFDG), which reflects quantitative glucose uptake into tissues and organs. However, data on the reproducibility of MRFDG measurements are scarce, particularly in patients with perturbed glucose homeostasis such as type 2 diabetes. We therefore aimed to evaluate the test-retest repeatability of both MRFDG and SUV in these patients.
Results:
Fifteen participants (mean age 71 ± 7 years; 2 females) with type 2 diabetes underwent a short 20-minute [18F]FDG D-WB PET/CT after 6 h fasting on two consecutive days. Both SUV and MRFDG images were reconstructed from D-WB PET/CT data obtained 60-80 min post-injection of [18F]FDG. MRFDG and SUV data were measured in organs and tissues, and repeatability was assessed with Bland-Altman analysis, intraclass correlation coefficients (ICC), repeatability coefficients (RPC) and coefficients of variation (wCV). There was high repeatability of both SUVmean and MRFDG-mean in all measured organs (ICC range: 0.65-0.95 for SUVmean and 0.66-0.94 for MRFDG-mean). SUVmean generally demonstrated higher reliability (ICC) and lower variability (%RPC and %wCV) when compared to MRFDG-mean. However, MRFDG test-retest variation was < 19% in most analysed tissues, demonstrating that MRFDG may be used as a precise marker of treatment response.
Conclusion:
This study demonstrates that MRFDG calculated from D-WB PET/CT exhibit high repeatability, comparable to SUVs across most organs in patients with type 2 diabetes.
More Related Videos
06:52Positron Emission Tomography Using 64-Copper as a Tracer for the Study of Copper-Related Disorders
Published on: April 28, 2023
08:07The Use of Ex Vivo Whole-organ Imaging and Quantitative Tissue Histology to Determine the Bio-distribution of Fluorescently Labeled Molecules
Published on: December 24, 2016
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...