Evaluation of visual function within the central 10 degrees using IMOvifa™ 24plus (1-2)
Yuki Takagi1,2, Ryo Asano1,3, Kana Yamashita2
1Chukyo Hospital 1-10, Nagoya, Aichi, Japan.
Plos One
|May 12, 2025
Summary
The IMOvifa™ perimeter shows strong correlation with the HFA 10-2 for visual field testing. While it may detect more residual visual function, it cannot replace the HFA 10-2 for central 10-degree evaluations.
Area of Science:
- Ophthalmology
- Visual Field Testing
- Medical Devices
Background:
- The central 10 degrees of the visual field are crucial for detailed assessment.
- The IMOvifa™ perimeter offers additional measurement points in this critical area.
- Comparative evaluation with existing perimetry methods is needed.
Purpose of the Study:
- To comparatively evaluate the IMOvifa™ 24plus (1-2) perimeter against the Humphrey Field Analyzer (HFA) 10-2.
- To assess the IMOvifa™ perimeter's ability to evaluate the central 10-degree visual field.
- To determine the concordance between IMOvifa™ 24plus (1-2) and HFA 10-2 for visual field defects.
Main Methods:
- 30 patients (48 eyes) underwent testing with IMOvifa™ 24plus (1-2), HFA 24-2, and HFA 10-2.
- Spearman's rank correlation was used for Mean Deviation (MD) and Pattern Standard Deviation (PSD).
- Kappa analysis evaluated concordance of visual field defects and residual visual field detection.
Main Results:
- IMOvifa™ 24plus (1-2) and HFA 24-2 showed strong correlation with HFA 10-2 for MD and PSD.
- Agreement with HFA 10-2 for detecting central 10-degree defects was substantial for both IMOvifa™ 24plus (1-2) (κ=0.715) and HFA 24-2 (κ=0.754).
- IMOvifa™ 24plus (1-2) detected residual visual function in 100% of cases where HFA 10-2 indicated it.
Conclusions:
- The IMOvifa™ 24plus (1-2) demonstrates potential for enhanced defect detection in specific visual field areas.
- It may also be effective in identifying residual visual function.
- However, the HFA 10-2 remains essential for comprehensive evaluation of the central 10-degree visual field.
More Related Videos
Related Concept Videos
Anatomy of the Eyeball
The eye is a spherical, hollow structure composed of three tissue layers. The outer layer — the fibrous tunic, comprises the sclera — a white structure — and the cornea, which is transparent. The sclera encompasses some of the ocular surface, most of which is not visible. However, the 'white of the eye' is distinctively visible in humans compared to other species. The cornea, a clear covering at the front of the eye, enables light penetration. The eye's middle layer, the vascular tunic,...
Photoreceptors and Visual Pathways
At the molecular level, visual signals trigger transformations in photopigment molecules, resulting in changes in the photoreceptor cell's membrane potential. The photon's energy level is denoted by its wavelength, with each specific wavelength of visible light associated with a distinct color. The spectral range of visible light, classified as electromagnetic radiation, spans from 380 to 720 nm. Electromagnetic radiation wavelengths exceeding 720 nm fall under the infrared category, whereas...