Diagnostic accuracy of an over-the-counter infant pulse oximeter for cardiorespiratory events
Colm P Travers1, Arie Nakhmani2, Kimberly M Armstead3
1Department of Pediatrics, The University of Alabama at Birmingham, Birmingham, Alabama, USA cptravers@uabmc.edu.
Objective:
To determine the diagnostic accuracy of an over-the-counter infant pulse oximeter for cardiorespiratory events.
Design:
Single-centre prospective diagnostic accuracy study.
Setting:
University of Alabama at Birmingham.
Patients:
Infants weighing ≥1500 g, <44 weeks' postmenstrual age (PMA) and off ventilator/continuous positive airway pressure support.
Interventions:
Test device for 48 hours in addition to standard hospital monitors, ECG and pulse oximetry.
Main Outcome Measures:
Data were time aligned and analysed using MATLAB. The coprimary outcomes were the diagnostic accuracy of the test device for the detection of events with heart rate (HR) <50 beats per minute (bpm) and events with oxygen saturations (SpO2) <80% for ≥3 s.
Results:
66 infants with a median gestational age of 31 weeks (range 23-40) were studied at a median 35 weeks' PMA (range 32-42) weighing 1930 g (range 1500-3605 g) from April to July 2023. The sensitivity for detection of HR <50 bpm ≥3 s was 6% and 39% for smoothed and raw data, respectively, while the specificity was >99% for both smoothed and raw data. The sensitivity for SpO2 <80% ≥3 s was 14% and 74%, while the specificity was >99% and 96% for smoothed and raw data, respectively. Sensitivity for bradycardia events was higher for events with longer durations and/or when using higher thresholds. Sensitivity was higher for hypoxaemia events with longer durations and/or when using higher thresholds.
Conclusion:
An over-the-counter infant pulse oximeter had high specificity for bradycardia and hypoxaemia events consistent with a low false alarm rate. Sensitivity improved with longer events and higher event thresholds.
Trial Registration Number:
NCT05774470.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: