Pär Wennberg1, Bengt Sundberg1, Elisabeth Westerdahl2
1School of Health Sciences, Jönköping University, Jönköping, Sweden.
Introduction:
A new flow-regulated PEP device has been evaluated regarding functionality and pressure properties.
Methods:
The three different resistance levels were assessed and evaluated at standardized flow rates of 10 and 18 L/min; Kruskal-Wallis test was used to analyse the differences in generated pressure between the different resistance levels.
Results:
A range of 3-31 cmH2O was generated with airflows of 10 and 18 L/min. There was a significant difference in pressure among different resistance levels at both flow rates.
Conclusion:
Overall, there was a significant difference in pressure among different resistance levels at both flow rates, showing that the high resistance significantly increased pressure compared with low resistance. This new device is performing comparable with other resistors available in the market.
More Related Videos
Related Concept Videos
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...