Related Experiment Video
Updated: May 5, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
S-ICD Implantation Following TV-ICD: Insights Into Patients With Infections and Abandoned Leads: The HONEST Cohort
Vasileios Sousonis1, Peggy Jacon2, Fawzi Kerkouri3
1Heart Rhythm Management Department, Clinique Pasteur, Toulouse, France.
Background:
Subcutaneous implantable cardioverter-defibrillators (S-ICDs) can be a viable option for patients with transvenous ICDs experiencing complications (rescue S-ICD).
Objectives:
This study sought to evaluate the outcomes of rescue S-ICD implantation using data from the HONEST French nationwide S-ICD cohort.
Methods:
All rescue S-ICD patients were identified. Outcomes (complications, reinterventions, and mortality) were compared between rescue and de novo S-ICD patients. Subgroup analyses were performed based on the implantation indication (infective vs noninfective) and the presence of abandoned leads.
Results:
Among 4,924 patients in the HONEST cohort, 651 underwent rescue S-ICD implantation (295 with infective indications, 244 with abandoned leads). Over a follow-up of 4.2 ± 2.2 years, complications and reinterventions were similar in rescue and de novo S-ICD patients (22.6% vs 21.0%; P = 0.35 and 8.7% vs 7.2%; P = 0.17, respectively), in infective and noninfective rescue S-ICD patients (21.6% vs 23.5%; P = 0.55 and 8.9% vs 8.6%; P = 0.90, respectively) and in patients with abandoned and extracted leads (24.3% vs 21.7%; P = 0.46 and 8.3% vs 9.0%; P = 0.77, respectively). Mortality was higher in rescue compared to de novo S-ICD patients (16.9% vs 10.2%; P < 0.001) and in rescue S-ICD patients with infective indications (29.2% vs 6.7%; P < 0.001) and extracted leads (21.9% vs 8.6%; P < 0.001), mainly due to a higher burden of comorbidities, as none of these parameters was independently associated with mortality in multivariate analyses. No S-ICD-related deaths were observed.
Conclusions:
Rescue S-ICD implantation is an acceptable option for patients with device-related complications. Lead abandonment in the setting of a noninfective indication appears to be safe. (S-ICD French Cohort Study [HONEST]; NCT05302115).
Related Concept Videos
Healthcare Associated Infections I: Iatrogenic, Exogenic and Endogenic
HAIs significantly increase the cost of health care. Extended stays in healthcare institutions, increased disability, increased costs of medications, including specialized antibiotics, and prolonged recovery times add to the patient's expenses and the healthcare institution and funding bodies.
Cardiac Catheterization IV: Nursing Management
Endocarditis III: Medical Management
Endocarditis IV: Nursing Management
Cardiomyopathy V: Interprofessional Care