Oscillation and Reaction Board Techniques for Estimating Inertial Properties of a Below-knee Prosthesis
Published on: May 8, 2014
Colorado Limb Donning-Timed Up and Go (COLD-TUG) Test in Lower-Extremity Amputation: Less Donning Time with
Mohamed E Awad1,2, Guy Lev2,3, Danielle H Melton2,3
1Department of Orthopedics, University of Colorado Anschutz Medical Campus, Aurora, Colorado.
The new Colorado Limb Donning-Timed Up and Go (COLD-TUG) test accurately measures prosthesis donning time for amputees. Bone-anchored limb (BAL) users demonstrated significantly faster donning times than traditional socket prosthesis users.
Area of Science:
- Prosthetics and Orthotics
- Rehabilitation Medicine
- Biomechanics
Background:
- Bone-anchored limb (BAL) prostheses offer direct skeletal integration for improved prosthesis attachment.
- Traditional Timed Up and Go (TUG) tests assess mobility but exclude prosthetic limb donning time.
- Assessing functional mobility in lower-extremity amputees requires incorporating donning time.
Purpose of the Study:
- To develop and validate the Colorado Limb Donning-Timed Up and Go (COLD-TUG) test.
- To assess the reliability and validity of the COLD-TUG test in amputees.
- To compare the COLD-TUG test performance between BAL and socket prosthesis users.
Main Methods:
- Participants with unilateral lower-extremity amputation were divided into socket prosthesis (n=15) and BAL (n=22) groups.
- The COLD-TUG test measured time for donning, standing, walking 3m, turning, returning, and sitting.
- Intrarater reliability and concurrent validity with the TUG test were analyzed.
Main Results:
- The COLD-TUG test demonstrated excellent intrarater reliability (ICC=0.94).
- Concurrent validity with the TUG test was good in BAL patients (r=0.712).
- BAL users had significantly shorter COLD-TUG times (16.6s) than socket users (85.3s).
Conclusions:
- The COLD-TUG test accurately quantifies prosthesis donning burden and functional mobility.
- BAL use is associated with reduced prosthesis donning time compared to socket prostheses.
- This test provides valuable insights into functional capacity and quality of life for amputees.
More Related Videos
11:16Engineering Platform and Experimental Protocol for Design and Evaluation of a Neurally-controlled Powered Transfemoral Prosthesis
Published on: July 22, 2014
07:52Evaluating Postural Control and Lower-extremity Muscle Activation in Individuals with Chronic Ankle Instability
Published on: September 18, 2020
Related Concept Videos
Bones of the Lower Limb: Tibia and Fibula
Peripheral Artery Disease V: Postoperative Nursing Management