Related Experiment Video
Updated: Jun 13, 2025

Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
Published on: July 15, 2009
Clinical Evidence on the Influence of Implant Position onto Maximum Output with the Bonebridge Bone Conduction
Daniel Schurzig, Rebecca Iseke1, Hannes Maier
1Department of Otorhinolaryngology, Hannover Medical School, Hannover, Germany.
Hypothesis:
In bone conduction implantation, the position of the implant influences the audiological benefit of the patient.
Background:
One way of treating hearing loss is the implantation of bone conduction implants (BCIs), which effectively transmit vibrations through the skull bone to the cochlea given that the implant transducer is securely fixated. Laboratory research on the efficacy of bone conduction sound transmission found that a closer proximity of the transducer to the ipsilateral cochlea yields significantly higher cochlear promontory vibrations and hence, higher stimulation efficacy. Up to now, this finding has not been reproduced using clinical data such as the functional or effective gain.
Methods:
The present, retrospective study was conducted on a cohort of 28 BCI patients to correlate the implantation site of the BC transducer, derived from clinical postoperative imaging and defined in a standardized coordinate system, with maximum output values that are exclusively based on a novel calculation method only employing clinical audiological data.
Results:
It could be shown that the efficacy of BCI stimulation is in fact correlated with the transducer distance to the cochlea, and that this correlation is frequency dependent. Furthermore, the longitudinal distance of the transducer and the ipsilateral external auditory canal is negatively correlated with the maximal output while the sagittal distance is not.
Conclusion:
The present study is hence the first one to clinically demonstrate the significance of BCI placement for maximizing patient benefit, which should be considered during the preoperative planning of bone conduction implantation.