Erythrocyte Sedimentation Rate Reference Intervals Determined via VES-MATIC 5 and CUBE 30 Touch with Respect to the
Maria Lorubbio1, Daniela Diamanti2, Carolina Pieroni2
1UOC Laboratory Medicine, Department of Laboratory Medicine and Transfusion, San Donato Hospital, 52100 Arezzo, Italy.
Diagnostics (Basel, Switzerland)
|May 14, 2025
Summary
New reference intervals for erythrocyte sedimentation rate (ESR) testing were established for automated analyzers, showing good agreement with the gold standard method. These findings support accurate clinical decision-making for inflammatory status assessment.
Area of Science:
- Clinical Chemistry
- Hematology
- Laboratory Medicine
Background:
- Erythrocyte sedimentation rate (ESR) is a crucial diagnostic test for assessing patient inflammatory status globally.
- Accurate reference intervals (RIs) are essential for interpreting ESR results and supporting clinical decisions.
- Existing RIs may not fully reflect performance of newer automated analyzers.
Purpose of the Study:
- To establish reference intervals (RIs) for the VES-MATIC 5 (VM5) and CUBE 30 touch (C30T) automated ESR analyzers.
- To compare the performance of VM5 and C30T against the gold standard ESR method.
Main Methods:
- A total of 989 presumably healthy participants were enrolled from Arezzo Hospital, Italy.
- ESR measurements were performed using the gold standard Westergren method, VM5, and C30T analyzers.
- Reference intervals were established according to Clinical Laboratory Standards Institute (CLSI) guidelines.
Main Results:
- Significant differences in ESR were observed between sexes and with increasing age across all methods.
- VM5 and C30T analyzers showed average and median values close to the gold standard.
- Specific RIs were determined for females (1-25 mm/h) and males across different age groups (1-18 mm/h and 1-29 mm/h).
Conclusions:
- The proposed reference intervals for the C30T and VM5 automated analyzers align with the gold standard method.
- These validated RIs can be reliably used for measuring ESR in EDTA blood samples.
- The study provides updated, method-specific RIs crucial for accurate inflammatory status assessment.