Concentration Estimation of Apixaban and Rivaroxaban in Plasma From Patients Through Heparin-Calibrated Anti-Xa Assay
Marta Elba Martinuzzo1, Claudio Sebastián Berger1, Emanuel Sueldo2
1Instituto Universitario del Hospital Italiano de Buenos Aires, CABA, Argentina.
International Journal of Laboratory Hematology
|May 14, 2025
Summary
Measuring apixaban (APIXA) or rivaroxaban (RIVA) drug levels using Heparin Calibrated Anti Xa (HC-Anti Xa) assays is feasible. However, results vary significantly between different reagent/calibrator/instrument systems, necessitating laboratory-specific calibration knowledge.
Area of Science:
- Clinical Chemistry
- Pharmacology
- Thrombosis Management
Background:
- Direct oral anti-Xa inhibitors (DOACs) like apixaban and rivaroxaban are crucial for treating and preventing thrombosis.
- While generally safe, measuring plasma drug concentrations is sometimes clinically necessary for these anticoagulants.
Purpose of the Study:
- To compare the performance of three different Heparin Calibrated Anti Xa (HC-Anti Xa) assay systems.
- To evaluate these systems for measuring apixaban (APIXA) and rivaroxaban (RIVA) concentrations in patient plasma samples.
Main Methods:
- Analyzed 49 apixaban and 40 rivaroxaban patient plasma samples using peak or trough concentrations.
- Measured drug concentrations using specific anti-Xa assays (Werfen, Stago) and compared results with three HC-Anti Xa systems (Werfen, Siemens, Stago).
Main Results:
- Significant differences in HC-Anti Xa values were observed across the three systems, with the Stago system yielding the highest results.
- High correlation (r² > 0.94) was found between Werfen and Siemens assays for both drugs, but Stago showed a plateau effect at high concentrations.
- Agreement improved for concentrations ≤ 200 ng/mL, but inter-system differences persisted at various clinical decision points.
Conclusions:
- Heparin Calibrated Anti Xa assays can estimate apixaban and rivaroxaban concentrations, but results are system-dependent.
- Laboratories must establish and understand the specific HC-Anti Xa values corresponding to critical anticoagulant concentrations for accurate clinical use.