Factors Influencing Discrepancies in Intraocular Pressure Measurements Between iCare IC200 and Goldmann Tonometer
Haowei Zhang1, Jeong Woo Nam, Mi Sun Sung
1Department of Ophthalmology and Research Institute of Medical Sciences, Chonnam National University Medical School and Hospital, Gwangju, South Korea.
PréCis:
Intraocular pressure measured by Goldmann applanation tonometry, central corneal thickness, and history of cataract surgery are factors in the discrepancies in intraocular pressure measurements between the iCare IC200 rebound tonometer and the Goldmann applanation tonometer.
Purpose:
To investigate the factors affecting the discrepancies in intraocular pressure (IOP) measurements between the iCare IC200 rebound tonometer (RBT) and the Goldmann applanation tonometer (GAT).
Materials And Methods:
The IOP was measured sequentially using RBT followed by GAT. Bland-Altman analysis was used to determine the agreement between the RBT and GAT. Eyes were classified into 3 groups based on the difference calculated as RBT minus GAT: group A (-1 to 1 mm Hg), group B (≥3 mm Hg), and group C (≤-3 mm Hg). Logistic regression analysis was performed to identify the factors associated with the measurement discrepancies.
Results:
A total of 402 right eyes were included in the study. The RBT and GAT measurements were significantly correlated (Spearman r =0.885, P <0.001). The mean IOP differences between the RBT and GAT were 0.114±2.13 mm Hg (range: -7 to 8, 95% limit of agreement: -4.06 to 4.29 mm Hg). Multivariate logistic analyses revealed that greater central corneal thickness (CCT) ( P =0.002) was significantly associated with overestimation of IOP by RBT relative to GAT. We found that underestimated IOP by RBT relative to GAT in patients with higher GAT-measured IOP ( P =0.002) and those with a history of cataract surgery ( P =0.006).
Conclusions:
RBT tends to overestimate IOP relative to GAT in patients with thicker CCTs, while it underestimates IOP in those with higher GAT-measured IOP or a history of cataract surgery.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Errors occurring during blood pressure monitoring
Several factors...