[Image Quality Assessment of Non-contrast Brain Computed Tomography Using High-pitch Double Spiral Scan: A Basic
Yuichiro Sato1, Tomomi Ohmura1, Shun Muramatsu2
1Department of Radiology and Nuclear Medicine, Akita Cerebrospinal and Cardiovascular Center.
Purpose:
Motion artifacts likely occur in patients with acute intracerebral hemorrhage (ICH). This study aimed to validate the image quality characteristics of non-contrast brain computed tomography (CT) with high-speed imaging technology, named Flash spiral (FS) (Siemens Healthineers, Erlangen, Germany). We verified the differences in CT values using a phantom that simulated hematoma.
Methods:
A dual-source CT scanner (SOMATOM Drive; Siemens Healthineers) was used to obtain reference and FS images of a Catphan700 phantom (The Phantom Laboratory, Greenwich, NY, USA). The CT values were measured in the hematoma-simulated acrylic module and urethane within the phantom. The noise power spectrum (NPS), task transfer function (TTF), and system performance function (SPF) between reference and FS images were obtained to compare image quality in each scan.
Results:
Compared with the phantom's reference and FS images, no significant differences were observed in the CT values between the samples simulating hematoma and their surrounding areas. The NPS showed lower values in the FS images than the reference images at spatial frequencies above approximately 0.4 cycles/mm, while the peak frequencies were nearly equivalent. The 10% TTF values were almost the same between both images. The SPF values were also equivalent between the two images at spatial frequencies above approximately 0.5 cycles/mm.
Conclusion:
In the phantom experiment, the 10% TTF values of the FS images were comparable to those of the reference images, indicating similar resolution in the high spatial frequency domain. FS is expected to expand the applicability for detecting cerebral hemorrhage in patients with significant body movement, where detection is challenging under standard conditions.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging