Evaluation of the performance of three pulse oximeters at different probe positions in awake rabbits
Giulia Maria De Benedictis1, Barbara Contiero1, Daniele Bovo1
1Department of Animal Medicine, Production and Health, University of Padova, Legnaro, Italy.
Abstract:
In rabbits, pulse oximetry plays a crucial role in monitoring oxygen saturation (SpO2) and pulse rate (PR), but it can be challenging due to their small size, variable anatomy, limited measurement sites, and potential motion artifacts during measurements. This prospective randomized study aimed to assess the reliability of three pulse oximeters on the forelimb and tail of fifty awake, healthy European rabbits. Two pulse oximeters, the Masimo Rad-5 (device 1) and the Edan VE-H100B, were tested, with the latter using both a Y-clip (device 2) and the Nellcor OxiMax adhesive finger sensor (device 3). Reliable values of SpO2 and PR were considered ≥ 95% and ≥ 151 bpm, respectively, which are normal values in a healthy rabbit. Success rates of the devices were calculated as the ratio of reliable to unreliable/missing values and compared using χ2 tests. Correlation and agreement between PR and clinically measured heart rate (cHR) were assessed using the Pearson correlation coefficient (r) and Bland-Altman analysis, respectively. Success rates differed significantly among devices (p < 0.001). The SpO2 success rates were 46%, 95.6%, and 94.4% for devices 1, 2, and 3, respectively. The PR success rates were 54.6%, 63%, and 72.8% for devices 1, 2, and 3, respectively. Success rates were higher on the forelimb (>66%) than on the tail (>26%) across devices. Correlation between PR and cHR was stronger on forelimb (r > 0.80) than on tail (r < 0.70). Agreement between PR and cHR was similar among devices, with a mean bias ranging from -6.6 to 6.7 bpm, and limits of agreement from -33.8 to 41.5 bpm. The Edan devices showed superior ability to detect reliable values of SpO2 and PR compared to the Masimo in rabbits. The forelimb appears to be a more reliable site for pulse oximetry in rabbits than the tail.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
09:20Novel Photoacoustic Microscopy and Optical Coherence Tomography Dual-modality Chorioretinal Imaging in Living Rabbit Eyes
Published on: February 8, 2018
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.