[Experimental evaluation of drainage properties of polymer stents under mechanical compression. Standard and tandem
E Mamaev I1,2, I Glinin K1,2, V Kotov S1,2
1FGAOU VO N.I. Pirogov Russian National Research Medical University of the Ministry of Health of Russia, Moscow, Russia.
Aim:
To assess the resistance to the mechanical compression of single and tandem stents.
Materials And Methods:
An in vitro model of the ureter using a silicone tube connecting the proximal and distal reservoirs was assembled. The proximal reservoir was sealed and connected to a pressure gauge. Stents were placed sequentially in the lumen of the tube. A part of the model of ureter was passed through a compression platform which had a length of 4 cm, a bending angle of 35-40 and the ability to apply variable pressure. Five different 7 Ch polymer ureteral stents were tested in single and tandem manner. The compression force was applied and measured at the level of the compression platform. A criterion of stent failure was an increase of the pressure in proximal reservoir over 20 cm H2O with irrigation of saline at a rate of 30 ml/hour.
Results:
Measurements were performed in 11 variants, including without a stent, with single stent (5) and with tandem stents (5). Polyurethane stents showed the least resistance to mechanical compression. In case of tandem stenting, all stents demonstrated an increase in resistance to compression compared to a single stent.
Conclusions:
Polymer stents are more resistant to mechanical compression compared to polyurethane ones. Tandem stenting can be an effective tool to counteract the compression by the tumor.
Related Concept Videos
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in ASTM C...
Testing Water Quality
Slump Test
Concrete is poured into the mold in three layers to conduct the test. Each layer is compacted 25 times with a steel tamping rod, which has a five-eighths-inch diameter and a rounded end, to ensure even distribution and eliminate...
Compacting Factor test
The procedure begins by placing concrete into the upper hopper without any compaction. Once filled, the bottom door of this hopper is opened,...
Behavior of Concrete Under Compressive Load
As the concrete specimen fractures under...