Introduction:
The objective of this study is to investigate and compare the impact of thyroid shields and lead glasses on the image quality (IQ) and artefacts in dental Cone Beam CT (CBCT) images. The current study aims to evaluate the effectiveness of this protective equipment in enhancing the quality of dental CBCT images.
Methods:
In this study, a New Tom dental CBCT device was used to scan Pro-Project phantom with and without thyroid shields and lead glasses in three different large, medium, and small Fields of View (FOVs). The several key IQ parameters such as noise, contrast-to-noise ratio (CNR), homogeneity, contrast-resolution (CR), and spatial-resolution (SR), were evaluated using ImageJ software. The measurements were performed and compared in various cross-sections based on the location of test objects in the phantom.
Results:
The study found that the use of thyroid shields and lead glasses resulted in a decrease in CNR and homogeneity parameters, while noise increased. However, there were no significant changes in SR and CR. It should be noted that the use of thyroid shields and lead glasses may increase image noise by 1-7 %. The combined use of lead shields was not included in our findings, as their effects are localized and similar to separate evaluations in adjacent areas in lower and upper sections.
Conclusion:
Thyroid shields and lead glasses during dental CBCT imaging increase image noise and reduce CNR. Proper placement is crucial, and reducing the FOV size can minimise artefacts while preserving diagnostic accuracy.
Implications For Practice:
Thyroid shields and lead glasses are vital for radiation protection. Increased noise demands attention in clinical practice. Ongoing research, training, quality assurance, and effective patient communication are essential for safety and trust in dental imaging.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...