Accuracy and Safety of a Continuous Noninvasive Blood Pressure Monitor in Neonates
R Brandon Hunter1,2, Michele S Saruwatari3, Nathan Daniel Bliss4
1Texas Children's Hospital, Houston, Texas, USA.
Introduction:
Accurate and continuous blood pressure (BP) monitoring in neonates is crucial in the intensive care unit. Invasive arterial lines (IALs), oscillometric cuffs, and current noninvasive continuous BP monitoring devices have significant limitations. The Boppli® device is a novel, continuous, noninvasive BP device that requires no calibration, designed for neonates.
Methods:
This prospective, multicenter study evaluated the performance, usability, and safety of the Boppli device in neonates <5 kg. We compared mean arterial pressure (MAP), systolic blood pressure (SBP), and diastolic blood pressure (DBP) measurements from the Boppli with IAL reference values by calculating average values of mean average error (MAE) and standard deviation (SD) for each patient, then averaging those means. Safety and usability were evaluated by analysis of adverse events and survey data, respectively.
Results:
The Boppli device demonstrated good performance, meeting the FDA requirements of MAE and SD of the entire cohort: MAE (SD) 0.7 (5.3) mm Hg for MAP, -0.8 (7.7) mm Hg for SBP, and 1.4 (4.7) mm Hg for DBP. Patients with elevated MAPs, Asian ethnicity, and lower extremity IALs were the subgroups with MAE >±5 mm Hg. Various subgroups had SDs >8 mm Hg attributed to low sample sizes. The device received high usability scores from clinicians and parents. No serious adverse events were reported.
Conclusion:
The Boppli device is a promising alternative for continuous noninvasive BP monitoring in neonates, offering good accuracy and usability. The device, which received 510(k) clearance in September 2023, was well received by clinicians and parents, with a low-risk profile.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...