Development and validation of a combined selected ion monitoring-scan GC-EI-MS method for nitazene analogs
Claire Phelps1, Emma K Hardwick1, Alleigh N Couch1
1Department of Forensic Science, Sam Houston State University, Huntsville, Texas, USA.
Abstract:
Novel synthetic opioids (NSOs) emerged during the third wave of the opioid epidemic, which has resulted in countless overdose deaths. Although fentanyl and fentanyl analogs are the most common NSOs, changing legislation in China and the United States led to the emergence of another class of NSOs, known as nitazene analogs. Nitazene analogs are very potent, and given their increasing prevalence in seized drug casework, it is imperative that seized drug laboratories have sufficient methods to identify nitazene analogs. This study developed a combined selected ion monitoring (SIM)-scan method to identify 20 nitazene analogs using gas chromatography-electron ionization-mass spectrometry (GC-EI-MS). The method was evaluated in terms of the limit of detection (LOD), carryover, selectivity, repeatability, reproducibility, and processed sample stability. The LOD for the targeted nitazene analogs ranged from 5 to 10 ppm due to increased sensitivity through SIM acquisition. No carryover was observed, and all nitazene analogs could be differentiated from other analogs and common interferences. The repeatability and reproducibility of the method were demonstrated qualitatively. All nitazene analogs were stable at room temperature for at least 24 hours. When applied to the analysis of blind simulant samples, 33 out of 35 samples were correctly identified, with the two misidentifications being due to ion ratios outside of tolerance. Likewise, two authentic nitazene analog casework samples were also correctly identified. The validated SIM-scan method provides a potential solution to identify low concentration nitazene analogs in seized drug casework that might otherwise be missed using routine GC-MS full scan acquisition.
More Related Videos
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
07:38A General Method for Detecting Nitrosamide Formation in the In Vitro Metabolism of Nitrosamines by Cytochrome P450s
Published on: September 25, 2017
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...