Individual and Simultaneous Imaging of 99 Tc and 177Lu with a Preclinical Broad Energy-Spectrum CZT-based SPECT
Pedro M C C Encarnação1, Pedro M M Correia1, Baharak Mehrdel2
1I3N - Physics Department, University of Aveiro, Portugal.
Abstract:
Radiopharmaceutical therapy has demonstrated a high efficacy in the treatment of various tumor types. One of the radionuclides already used in the clinic is 177Lu, a beta emitter that also emits several photons imageable with SPECT. Quantitative imaging of 177Lu is critical for developing new radiopharmaceuticals. Energy resolution is an important factor when imaging multiple photon emissions. Solid-state detectors offer a superior performance over scintillators, that are commonly used in commercially-available preclinical SPECT scanners. This study demonstrates the feasibility of 99 Tc and 177Lu quantitative imaging in mouse phantoms, individually and simultaneously, with a SPECT prototype built with four CdZnTe (CZT) detector heads and a custom-designed and energy-optimized parallel-hole tungsten collimator. With a custom implementation of the one-step late (OSL) image reconstruction algorithm, the system is capable of imaging energies from ~70 keV to 250 keV. Above 250 keV, images were significantly affected by septal penetration, consistent with the collimator design. A recovery coefficient within 25% was obtained for activities as low as 2 kBq/mL for 99 Tc and 45% for 177Lu. Compared to a commercial NaI-based preclinical SPECT (VECTor4/CT), our prototype showed a superior energy resolution (< 5% at 140 keV), a similar uniformity with a high-compact design.
More Related Videos
16:10A TIRF Microscopy Technique for Real-time, Simultaneous Imaging of the TCR and its Associated Signaling Proteins
Published on: March 22, 2012
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Overview of Microscopy Techniques