Related Experiment Video
Updated: Sep 23, 2026

Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Valve-in-valve transcatheter aortic valve replacement (TAVR) leads to lower device success compared to TAVR in native
Michael Paukovitsch1, Bartu Dilaver1, Dominik Felbel1
1Department of Cardiology, Ulm University Heart Center, Ulm, Germany.
Background:
Despite the lack of randomized-controlled trials in patients with failed bioprosthetic valves, valve-in-valve transcatheter aortic valve replacement (ViV-TAVR) is increasingly used.
Methods:
Outcomes of consecutive patients treated with ViV-TAVR (N = 100) at our tertiary heart center between 2014 and 2022 were compared to TAVR (N = 2216) in native valves.
Results:
Patients median age was 78.5 years (IQR 70.0-84.0) in ViV-TAVR compared to 81.0 (IQR 77.0-85.0) in patients with native aortic stenosis (p < 0.01) with a similar percentage of females in both groups (42% vs. 49.3%, p = 0.18). The median Society of Thoracic Surgeons score for mortality was significantly higher in patients undergoing ViV-TAVR [5.1% {IQR 2.6%-8.6%} vs. 3.8% {IQR 2.4%-6.3%}, p < 0.01]. ViV-TAVR was performed in degenerated surgical bioprostheses in 88% and in degenerated transcatheter bioprostheses in 12%. Stenosis was the main mechanism of bioprosthetic valve failure (70%), whereas severe regurgitation was the leading cause in 30%. The overall rate of device success amounted to 66% in ViV-TAVR, compared to 96.1% in TAVR (p < 0.01) and ViV-TAVR was independently associated with reduced device success (OR: 0.07, 95%CI: 0.045-0.12, p < 0.01) in multivariate regression. While ViV-TAVR decreased peak and mean gradients significantly, in 31% of patients elevated mean gradients (≥20 mmHg) were observed at discharge. Small native prosthesis diameter (<20 mm) was the strongest predictor (OR 3.8, 95%CI: 1.5-9.2, p = 0.01) independently associated with elevated gradients after ViV-TAVR.
Conclusion:
ViV-TAVR for treatment of degenerated bioprostheses improves aortic valve function. However, device success is lower compared to TAVR in native aortic valve disease, mainly due to elevated postprocedural mean gradients, especially in small bioprostheses.
More Related Videos
08:50Technique and Patient Selection Criteria of Right Anterior Mini-Thoracotomy for Minimal Access Aortic Valve Replacement
Published on: March 26, 2018
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Mitral Stenosis III: Medical Management
Aortic Regurgitation III: Medical Management