Is invasive fractional flow measurement accurate in intracranial stenosis? A computational simulation study
Yu Liu1, Zhengzheng Yan2, Ziqi Li1
1Department of Medicine and Therapeutics, The Chinese University of Hong Kong, Hong Kong, China.
Background:
In intracranial atherosclerotic stenosis (ICAS), low fractional flow (FF) may indicate hemodynamic significance.
Objective:
To investigate, using simulation models, whether invasive measurement could accurately reflect 'true' FF, when the catheter/pressure wire might disturb focal flow during measurement.
Methods:
We recruited 5 patients with high-grade, symptomatic M1 middle cerebral artery stenosis in three-dimensional rotational angiography (3DRA). In each case, the stenotic severity was manually manipulated to 50%, 60%, 70%, and 80%. At each stenotic severity, we simulated four situations: no catheter/pressure wire insertion ('in vivo' status); only catheter inserted proximally to the ICAS; catheter inserted and pressure sensor placed at 1 cm and 2 cm distally to the ICAS lesion. The blood flow was simulated with computational fluid dynamics modeling, and FF measured as post-stenotic and pre-stenotic pressure ratio. We calculated relative differences of FFs simulated at other situations compared with in vivo status.
Results:
Compared with in vivo status, catheter insertion had no significant influence on simulated FFs. With pressure wire passing through the ICAS, simulated FFs slightly decreased (mostly <10%) at 50% and 60% stenoses, which significantly decreased (up to 50% and 88%) at 70% and 80% stenoses. The effects of pressure wire on FFs were similar when the pressure sensor was placed at 1 cm and 2 cm distally to the ICAS.
Conclusions:
Invasive measurement of FF may overestimate the hemodynamic significance of ICAS with severe stenosis, as the pressure wire may further reduce the flow across the small residual lumen. The findings warrant verification in larger-scale studies, with information on collateral circulation and validation with other imaging modalities.
More Related Videos
13:07Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
11:16In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
Related Concept Videos
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
Pipe Flowrate Measurement
The orifice meter is a simple,...