Evaluation of real-world application of cardiac implantable electronic device-based multi-sensor algorithm for heart
Jennifer Llewellyn1, Rachel Goode1, Matthew Kahn1
1Department of Cardiology, Liverpool Heart and Chest Hospital, Liverpool L14 3PE, UK.
Aims:
Remote monitoring of cardiac implantable electronic devices enables pre-emptive management of heart failure (HF) without additional patient engagement. The HeartLogic™ algorithm in implantable cardioverter defibrillators (ICDs) combines physiological parameters to predict HF events, facilitating earlier interventions. This study evaluated its diagnostic performance and resource implications within an HF management service.
Methods And Results:
In a single-centre study, 212 patients with cardiac resynchronization therapy ICDs (CRT-Ds) were monitored for 12-months. During follow-up, 18 (8%) patients died, and 15 HF hospitalizations occurred in 13 (6%) patients. Outpatient visits totalled 37 in 34 (16%) patients. HeartLogic™ alerts occurred in 58% of patients, with 100% sensitivity for HF-related hospitalizations. The positive predictive value was 29% including only alerts associated with HF events, while it was 51% including HF events and explained alerts. Unexplained alert rate was 0.46 per patient-year. Clinical interventions, mainly medication adjustments, followed 82 alerts. Total management time was 257 h/year, equivalent to 0.57 full-time equivalents for managing 1000 CRT-D patients.
Conclusion:
The integration of HeartLogic™ into routine care demonstrated its utility in optimizing HF management, improving healthcare resource allocation. The algorithm can enhance proactive patient management and provide holistic care within the existing healthcare infrastructure.
More Related Videos
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pathophysiology of Heart Failure
Heart Failure Drugs: Inhibitors of Renin-Angiotensin System