Evaluating needle-free connectors associated backflow in Midline and peripherally inserted central catheters: A top
Daniele Privitera1,2, Davide Giustivi3, Elisa Nardin1
1Department of Cardiovascular, Neural and Metabolic Sciences, Istituto Auxologico Italiano, IRCCS, San Luca Hospital, Milan, Italy.
The Journal of Vascular Access
|May 22, 2025
Summary
Needle-free connectors (NFCs) reduce needlestick injuries but can cause backflow. Positive displacement NFCs showed lower backflow volume with peripherally inserted central catheters (PICCs) compared to other types.
Area of Science:
- Vascular access devices
- Medical device safety
- Fluid dynamics in catheters
Background:
- Needle-free connectors (NFCs) are crucial for reducing needlestick injuries in vascular catheterization.
- NFCs are categorized by their fluid displacement: positive, negative, neutral, and anti-reflux.
- Different NFC types and clamping sequences may influence fluid dynamics within catheters.
Purpose of the Study:
- To evaluate the backflow volume (BV) in vascular catheters using different NFC categories.
- To compare BV across NFC types when used with peripherally inserted central catheters (PICCs) and Midline catheters (MCs).
- To assess the impact of clamping sequences on BV with various NFCs.
Main Methods:
- A bench study evaluated four NFC types with 4 Fr x 60 cm PICC and 4 Fr x 25 cm MC.
- The experimental setup simulated physiological superior vena cava blood pressure.
- Three operators performed standardized sealing sequences, and subsequent BV was measured in mm³.
Main Results:
- No NFC completely eliminated backflow into the catheter.
- Positive displacement NFCs demonstrated significantly lower BV with PICCs (0.83 mm³) compared to other NFCs.
- No significant differences in BV were observed between NFC types when used with MCs.
- Clamping sequences did not affect BV for neutral and anti-reflux NFCs.
Conclusions:
- While no NFC prevents backflow entirely, positive displacement types offer reduced reflux with PICCs.
- The choice of NFC may be critical in minimizing backflow, particularly with PICCs.
- Further research could explore optimal NFC selection and usage protocols for different catheter types.