International Council for Standardization in Haematology (ICSH) Guidance on the Validation of Laboratory Developed
Chris Gardiner1, Akbar Dorgalaleh2, Marión Echenagucia3
1St James's University Hospital, Leeds, UK.
Laboratory developed tests (LDTs) are widely used in clinical hemostasis laboratories. An LDT may be defined as an in vitro diagnostic (IVD) test that is designed, manufactured, and used within a single laboratory. As with all other clinical laboratory tests, LDTs must be validated to ensure fitness for purpose. This may include the assessment of accuracy/comparability, precision, analytical sensitivity, and specificity, reportable range, reference intervals, linearity, and carryover. Not all validation elements will be applicable to all situations, and this will be dictated by the type of assay, the intended use, and the laboratory setting, for example a minor modification of an assay with regulatory approval will require fewer validation procedures than a wholly new test using reagents developed within the laboratory. Many LDTs in the hemostasis laboratory cannot be assessed in the usual fashion, so alternative approaches must be developed.
Laboratory developed tests (LDTs) are widely used in clinical hemostasis laboratories. An LDT may be defined as an in vitro diagnostic (IVD) test that is designed, manufactured, and used within a single laboratory. As with all other clinical laboratory tests, LDTs must be validated to ensure fitness for purpose. This may include the assessment of accuracy/comparability, precision, analytical sensitivity, and specificity, reportable range, reference intervals, linearity, and carryover. Not all validation elements will be applicable to all situations, and this will be dictated by the type of assay, the intended use, and the laboratory setting, for example a minor modification of an assay with regulatory approval will require fewer validation procedures than a wholly new test using reagents developed within the laboratory. Many LDTs in the hemostasis laboratory cannot be assessed in the usual fashion, so alternative approaches must be developed.
Related Concept Videos
Regulation of Hematopoietic Stem Cells
Immunocytochemistry and Immunohistochemistry
These...
Introduction to Hemostasis
The three phases of hemostasis involve many clotting factors present in plasma and several substances released by platelets and injured tissue cells. It is a fast, localized,...