Related Experiment Video
Updated: Sep 20, 2025

A Simplified Stepwise Approach to Echo Guidance during Percutaneous Mitral Valve Repair
Published on: October 16, 2021
Valve Underexpansion and Clinical Outcomes With ACURATE neo2: Findings From the ACURATE IDE Trial
Raj R Makkar1, Tarun Chakravarty1, Aakriti Gupta1
1Cedars-Sinai Medical Center, Los Angeles, California, USA.
Background:
In the ACURATE IDE (Safety and Effectiveness Study of ACURATE Valve for Transcatheter Aortic Valve Replacement) randomized controlled trial, ACURATE neo2 failed to show noninferiority to commercially available balloon-expandable (SAPIEN 3/3 Ultra) and self-expanding (Evolut R/PRO/PRO+/FX) valves for the primary endpoint of all-cause mortality, stroke, or rehospitalization at 1 year. A retrospective investigation was undertaken to evaluate potential factors contributing to these outcomes.
Objectives:
The goal of this study was to assess the impact of ACURATE neo2 valve expansion on clinical outcomes in the ACURATE IDE trial.
Methods:
Post hoc case review identified angulated (nonparallel) commissure posts in a few implanted ACURATE neo2 valves, indicating valve underexpansion. Procedural angiograms for all ACURATE neo2 valves implanted in the trial's main randomized cohort (n = 752) were inspected by an independent core laboratory. An exploratory analysis was performed to evaluate the association between valve expansion and clinical outcomes.
Results:
Of the 624 patients who underwent implantation with the ACURATE neo2 and had evaluable procedural angiograms, 135 (21.6%) had underexpanded valves. Greater aortic valve leaflet and annulus calcification at baseline was independently associated with ACURATE neo2 valve underexpansion (OR: 1.92; 95% CI: 1.27-2.91; P = 0.002). Procedural techniques, including frequency of predilation (100% in both groups) and postdilation (26.7% vs 25.2%; P = 0.72), and balloon sizing did not differ between the underexpanded and expanded valve groups. ACURATE neo2 underexpansion was associated with a higher 1-year rate of death, stroke, or rehospitalization (underexpanded: 18.7%; expanded: 11.8%; P = 0.04), which was confirmed in a multivariable analysis (HR: 1.92; 95% CI: 1.27-2.91; P = 0.002).
Conclusions:
Underexpansion of the ACURATE neo2 valve in the ACURATE IDE study was associated with a higher risk of the composite endpoint of death, stroke, or rehospitalization. Given the post hoc nature of these analyses, the study findings should be considered hypothesis generating. Whether achieving optimal valve expansion of the ACURATE neo2 valve with improvement in device design and procedural iterations will translate into improved clinical outcomes remains to be studied.
More Related Videos
06:56Author Spotlight: Advancing Cardiac Procedure Testing Prior to Embarking on Large Animal Studies
Published on: August 25, 2023
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Mitral Stenosis II: Clinical features and Diagnostic Tests
Mitral Regurgitation II: Clinical Features and Diagnostic Tests
Acute Coronary Syndrome III: Diagnostic Studies
Aortic Regurgitation III: Medical Management
Aneurysm II: Clinical Manifestations and Diagnostic Studies